Title: | Inter-comparison of absorbed dose to water in a Co-60 therapy beam using IAEA and HPA protocols |
Authors: | Farai, IP Kadni, T |
Keywords: | Absorbed dose Teletherapy Co-60 beam quality Dosimeter |
Issue Date: | 2000 |
Publisher: | COLLEGE OF MEDICINE, UNIVERSITY OF IBADAN, NIGERIA |
Citation: | Afr. J. Med. Med. Sci. (2000) 29, 219 -221 |
Abstract: | The absorbed dos e water in a Co-60 teletherapy beam has been measured with four different standard ionisation chambers applying two codes of practice and also, with a Fricke dosimeter. Measured values agree generally within 2.3%. There is a remarkable agreement of less than 0.3% variation, between ionisation chambers NE 2561 and NE 2481 when the IAEA protocol is applied. The HPA protocol is applicable to only NE 2561 and a variation of about 1.4% was observed between measurements made with this ionisation chamber applying the two protocols. The IAEA protocol shows very accurate results. With a deviation of about 2.2%, the Frickle dosimeter appears to be the least accurate for therapy dose measurement despite its simplicity of application. |
Description: | Article |
URI: | http://adhlui.com.ui.edu.ng/jspui/handle/123456789/3823 |
ISSN: | 1116-4077 |
Appears in Collections: | African Journal of Medicine and Medical Sciences |