Title: Inter-comparison of absorbed dose to water in a Co-60 therapy beam using IAEA and HPA protocols
Authors: Farai, IP
Kadni, T
Keywords: Absorbed dose
Teletherapy
Co-60 beam quality
Dosimeter
Issue Date: 2000
Publisher: COLLEGE OF MEDICINE, UNIVERSITY OF IBADAN, NIGERIA
Citation: Afr. J. Med. Med. Sci. (2000) 29, 219 -221
Abstract: The absorbed dos e water in a Co-60 teletherapy beam has been measured with four different standard ionisation chambers applying two codes of practice and also, with a Fricke dosimeter. Measured values agree generally within 2.3%. There is a remarkable agreement of less than 0.3% variation, between ionisation chambers NE 2561 and NE 2481 when the IAEA protocol is applied. The HPA protocol is applicable to only NE 2561 and a variation of about 1.4% was observed between measurements made with this ionisation chamber applying the two protocols. The IAEA protocol shows very accurate results. With a deviation of about 2.2%, the Frickle dosimeter appears to be the least accurate for therapy dose measurement despite its simplicity of application.